IM-106-5500, Original Issue
August 2005
CCO 5500
4-16
Serial Input — If this option is selected the criteria controlling plant status 
are transmitted via the serial data link. Select this option by pressing 
ENTER when the 'Plant Status - Serial' screen is displayed. 
Multiple  Four options are available here. Press ENTER when the 'Plant 
Status - Multiple' screen is displayed. The first option is displayed - 
Temperature. Use the arrow key to toggle YES or NO. NO will mean that 
the temperature threshold is not used to determine plant status. If YES is 
selected the display will enter the display below. Configure the instrument 
for temperature threshold, press ENTER when correctly configured and 
the display will move to the next option. Select YES or NO in a similar 
manner to above. After the last option has been set (Logic Input) the 
display will return to the 'PARAMETERS - Plant Status I/P' screen. Use the 
down arrow to scroll down to EXIT and press ENTER. The plant status is 
now fully configured. 
Plant status will only be OFF if all options selected are registering plant 
status OFF. If any one of them is not fulfilling plant status OFF conditions, 
then the instrument will register plant status ON. 
Temperature Threshold — A value is set here for the temperature 
threshold. While the temperature (taken from the normalizing 
temperature) is above the threshold value, plant status is ON. If the 
temperature drops below the threshold, plant status is OFF and only 
the seconds averaging stack will update. 
Oxygen Threshold — Set and used in a similar manner to the 
temperature threshold. However, if the normalizing oxygen level rises 
above the threshold, plant status is OFF. For plant status ON, the 
oxygen level must be below the threshold set. 
Water Vapor — Set and used in a similar manner to the temperature 
threshold. If the normalizing water vapor level falls below the threshold, 
plant status is OFF. For plant status ON, the water vapor level must be 
above the threshold set. 
Logic Input — Select YES or NO and press ENTER. For plant status 
to be ON the logic input (PS1 and PS2) must be open circuit; for plant 
status to be OFF the logic input must be closed circuit. After this option 
has been configured the menu will exit to 'Multiple'. Use the down 
arrow to select EXIT and press ENTER.