TEST POINTS
If further checks on instrument operation are required, there are various test 
points within the instrument. Many of these are simple DC voltages and so 
may be checked using a voltmeter set to DC volts, otherwise may be 
investigated using an oscilloscope: 
Signal processor
T1
D1 signal from receiver head with some 
conditioning in the processor, smoothed by a 
factor of divide by 10. Flattened saw-tooth, 32 
to 45 Hz (modulation frequency) 1V pk-pk 
(maximum), centred on 0V. 0V for T1 to T6 
may be taken from 0V test point, top left of 
board. Test-points T1 to T6 are to be found in 
the center of the board.
T2
D1 signal after amplification in processor. 
Flattened saw-tooth, 32 to 45 Hz, 3.5V pk-pk 
(maximum), centered on 0V. Amplitude may be 
varied by a trim pot. Refer to Section 4, 
Detector Levels.
T3
D1 output to the A/D converter within the 
micro-processor, offset by 2.5V. Flattened 
saw-tooth, 32 to 45 Hz, 3.5V pk-pk (maximum), 
centered on 2.5V. 
T4
As T1 but for D2 signal.
T5
As T2 but for D2 signal.
T6
As T3 but for D2 signal.
T7
Receiver ambient temperature to the A/D 
converter with the micro-processor - 1mV 
represents 33.8
o
F (1
o
C).
T8
Normalizing input for pressure, before A/D 
converter 0.8 to 4.0V = 4 to 20 mA.
T9
As T8 except oxygen.
T10
As T8 except temperature.
+5 and 0V
Supply rails for the micro-processor. DC 
voltage.
12V
Supply rail for the plant status input.
+15V, -15V and 0VA
Supply rails for the analog current output.
VIN
Output from the D/A converter. 0 to 2.5V 
represents 0 to 20 mA (4 mA at 0.V).
-15V1, 0V and +15V1
Isolated supply for the analog current output.
0VB and +12VB
Isolated supply for the plant status input.
F
Reference wave from the transmitter unit (via 
the power supply). Square wave, 30 top 45 Hz, 
5V pk-pk, centered on 2.5V.
Power Supply
0V and +12V
Power supply for the instrument.
Receiver
0V
0V for the receiver.
T1
Detector output without conditioning.
T2
Detector output after first stage of gain.
T3
Detector output after both stages of gain.
Transmitter
S-
0V supply to the heater cartridge.
S+
12V supply to the heater cartridge.
M+ and M-
Supply to the chopper motor (+1V DC).
T3
Reference wave.
T4
Reference wave.