HP (Hewlett-Packard) HP 8360 Manuel D’Utilisation

Page de 508
 COMMAND SUMMARY
Pulse Subsystem
Since frequency and period are inversely related, if both are sent
in the same message, only the last one is applied. If the 
command and either the 
 or 
 command are sent
in the same message, they must be accepted without error if the
resulting pulse is possible.
PULSe:FREQuency 
 
PULSe:FREQuency ?
 
Sets and queries the frequency of the internal pulse generator. The
resolution of the frequency is such that the resulting period is set to a
resolution of 1 
 The *RST value is 500 
PULSe:PERiod 
 suffix] 
PULSe:PERiod? 
Sets and queries the period of the internal pulse generator. The
resolution of this is 1 
 The *RST value is 2 
PULSe:WIDTh 
 
PULSe:WIDTh? 
Sets and queries the width of the internal pulse generator. The *RST
value is 1 
PULM:SLEW 
 
PULM:SLEW? 
Sets and queries the rise time for the pulse modulation. The typical
usage is MAX I MIN since calibrating the rise time of the pulses is not
common. Slow pulse is set by the command PULS:SLEW MAX. Any
value above 1.8 
 is set to maximum. The *RST setting is 
MIN.
 
PULM:SLEW:AUTO?
Sets and queries the automatic setting of rise time for the pulse
modulation system. The *RST setting is ON.
PULM:SOURce 
PULM:SOURce?
Sets and queries the source for the pulse modulation control signal.
*RST value is 
PULM:STATe 
 PULM:STATe?
Sets and queries the state of pulse modulation. The *RST value is 0.
ROSCillator:SOURce?
ROSCillator:SOURce 
Sets and queries the reference oscillator selection switch. The
HP 8360
User’s Handbook
Operating and Programming Reference