Emerson Series 3000 MVD Manuel D’Utilisation

Page de 348
Configuration and Use Manual
103
Configuring Digital Communications
Co
nfi
guri
ng Custod
T
ran
sf
er
Star
tup
Con
fig
urin
g Tic
kets
Con
fig
ur
in
g Di
gi
tal
 Co
mm
u
n
ic
a
tio
ns
13.4.1
Loop current mode
Loop current mode is used to fix or unfix the mA output:
If Loop Current Mode is disabled, the mA output is fixed at 4 mA, and therefore cannot be 
used to report process data.
If Loop Current Mode is enabled, the mA output will report process data as configured.
13.4.2
Burst mode
Burst mode (or HART burst mode) is a specialized mode of communication during which the 
Series 3000 device regularly broadcasts process variable data over the primary mA output, using 
HART protocol. The Burst Command parameter allows you to specify the content and format of the 
process variable data.
Burst mode is ordinarily disabled, and should be enabled only if another device on the network 
requires HART burst mode communication.
Burst command
Transmitter vars
Specify the type of information that will be broadcast via burst mode. Select 
from:
Primary variable – The Series 3000 device repeats the primary variable 
(PV)
(4)
, in measurement units, in each burst (e.g., 14.0 g/s, 13.5 g/s, 12.0 g/s).
PV & pct range – The device sends the PV’s percent of range and the PV’s 
actual mA level in each burst (e.g., 25%, 11.0 mA).
Dyn vars & Pct range – The device sends PV, SV, TV, and quaternary variable 
(QV) values in measurement units and the PV’s actual milliamp reading in 
each burst (e.g., 50 lb/min, 23 °C, 50 lb/min, 0.0023 g/cm
3
, 11.8 mA).
Transmitter vars – In each burst, the device sends the four process variables 
configured for burst variables 1–4.
Burst variable 1
(5)
Mass flow rate
• Specifies the process variable to be broadcast in position 1. 
• Select the desired process variable from the list.
Burst variable 2
Temperature
• Specifies the process variable to be broadcast in position 2. 
• Select the desired process variable from the list.
Burst variable 3
Density
• Specifies the process variable to be broadcast in position 3.
• Select the desired process variable from the list.
Burst variable 4
Mass total
• Specifies the process variable to be broadcast in position 4.
• Select the desired process variable from the list.
(1) If HART protocol is configured for RS-485, Polling Address in RS-485 menu and Polling Address in Bell 202 menu are duplicates. 
(2) See Section 13.4.1.
(3) See Section 13.4.2.
(4) The primary variable (PV) is the process variable reported via the primary mA output (MAO1). The secondary variable (SV) is the 
process variable reported via the secondary mA output (MAO2). The tertiary variable (TV) is the process variable reported via the 
frequency output. The quaternary variable (QV) is accessible via HART, and can be assigned through the Device Setup menu. See 
Section 13.5.
(5) This parameter appears only if Transmitter Vars is specified for Burst Command.
Table 13-4
Bell 202 parameters continued
Variable
Default
Description