Emerson Series 3000 MVD Manuel D’Utilisation

Page de 348
140
Micro Motion
®
 Series 3000 MVD Transmitters and Controllers
Startup Procedures
Milliamp outputs
Refer to Figure 16-2 to access and set the output levels of milliamp outputs.
1. Select the milliamp output to be set.
2. Use the cursor control buttons to change the output value.
3. Press 
SAVE
 to set the output level.
4. Check the receiving device to ensure that the correct output level is being received.
When you exit to the Simulate Outputs screen, the milliamp output is released and is again controlled 
by the application.
If the difference between the output level sent by the Series 3000 device and the output level received 
by the receiving device is not acceptable for your application, perform a milliamp output trim. See 
Section 16.5.
Frequency output
Refer to Figure 16-2 to access and set the output levels of the frequency output.
1. Select the frequency output.
2. Use the cursor control buttons to change the output value.
3. Press 
SAVE
 to set the value.
4. Check the receiving device to ensure that the correct output level is being received.
When you exit to the Simulate Outputs screen, the frequency output is released and is again controlled 
by the application.
16.5
Milliamp output trim
milliamp output trim creates a common measurement range between the Series 3000 device and the 
device that receives the mA output. For instance, a Series 3000 device might send a 4 mA signal that 
the receiving device reports incorrectly as 3.8 mA. If the Series 3000 device output is trimmed 
correctly, it will send a signal appropriately compensated to ensure that the receiving device actually 
indicates a 4 mA signal.
You must trim the outputs at both the 4 mA and 20 mA points to ensure appropriate compensation 
across the entire range of outputs.
To perform a milliamp output trim:
1. Connect a digital multimeter (DMM) or other reference device to the primary or secondary 
milliamp wiring terminals. Ensure that the reference device is connected in series. Table 16-1 
lists the terminals to which the reference device should be connected. 
2. Ensure that the reference device is set to read milliamps.