Operation Mode
17.5.6
Diagnostic monitor
The Diagnostic Monitor shows real-time values for sensor flow tube frequency, left and right pickoffs, 
drive gain, and live zero.
Drive gain, pickoff levels, and sensor flow tube frequency values are useful for troubleshooting fault 
alarms. For information about troubleshooting fault alarms, see Chapter 22.
Live zero is useful for monitoring the flow rate when it drops below the mass low flow cutoff. To 
configure the mass low flow cutoff, see Section 7.3.2.
Note: The diagnostic monitor is not the same as the Diagnostics menu. The Diagnostics menu, 
accessed through the Maintenance option of the Management menu, allows you to read the values of 
external inputs and simulate output levels for outputs. The Diagnostics menu and its use are 
described in Chapter 16.
17.5.7
LCD options
The LCD Options allow you to adjust the contrast, or to enable screen backlighting. 
1. Select 
LCD Options
 from the View menu.
2. Select 
Contrast
 to adjust the screen contrast.
3. Select 
LCD Backlight
 to turn screen backlighting on or off.
17.5.8
Density curves
The Density Curves menu item is available only if the enhanced density application is installed and 
one or more density curves has been configured. This screen lists all density curves that have been 
configured. The active density curve is highlighted.
If no density curve has been assigned to a batch preset, you can use this menu to change the 
active density curve. To do this, highlight the desired curve in the list and press 
SEL
.
If a density curve has been assigned to a batch preset, that curve is active and you cannot 
change the active density curve from this menu. 
17.5.9
Applications list
The Applications List shows the software version of the Series 3000 device and the core processor (if 
applicable) and all applications that are installed. Refer to this screen for the software revision number 
to report problems.