Emerson Series 3000 MVD Manuel D’Utilisation

Page de 348
Configuration and Use Manual
181
Measurement Performance
Dia
g
nostics and T
roub
leshooting
Spe
c
ifica
tions – Model 
3
350
/3
7
0
0
Specific
a
tions
 –
 Model 
330
0
/3
500
Measure
ment P
e
rf
or
mance
21.2.2
Meter validation and meter factors
Meter validation can be used with all Series 3000 systems (both transmitters and controllers). There 
are no special requirements.
Meter validation compares a measurement value reported by the Series 3000 device with an external 
measurement standard. Meter validation requires one data point.
Note: For meter validation to be useful, the external measurement standard must be more accurate 
than the sensor. See the sensor’s product data sheet for its accuracy specification.
If the Series 3000 device’s mass flow, volume flow, or density measurement is significantly different 
from the external measurement standard, you may want to adjust the corresponding meter factor. A 
meter factor is the value by which the Series 3000 device multiplies the process variable value. The 
default meter factors are 
1.0
, resulting in no difference between the data retrieved from the sensor and 
the data reported externally.
Meter factors are typically used for proving the flowmeter against a weights and measures standard. 
You may need to calculate and adjust meter factors periodically to comply with regulations.
21.2.3
Calibration
Calibration applies only to the Model 3500 and Model 3700 transmitters. There are no special 
requirements.
Result data on display
Pass/Fail/Abort for current test
For all results stored on transmitter: 
• Pass/Fail/Abort
• Abort code (if relevant)
• Stiffness of the right and left pickoffs
Result data with Communicator Pass/Caution/Abort for current test
For all results stored on transmitter: 
• Pass/Caution/Abort
• Abort code (if relevant)
• Stiffness of the right and left pickoffs
• Comparison table for stored results
• Comparison plot for stored results
Result data with ProLink II
For all results stored on PC:
• Pass/Fail/Abort
• Abort code (if relevant)
• Stiffness of the right and left pickoffs
• Test execution metadata
• Comparison graphs
• Test reports
• Data export and manipulation 
capabilities
For all results stored on transmitter:
• Pass/Fail/Abort
• Abort code (if relevant)
• Stiffness of the right and left pickoffs
• Test execution metadata
• Comparison graphs
• Test  reports
• Data export and manipulation 
capabilities
Startup methods
Manual
Manual
Scheduler
Event
Discrete input
Table 21-2
Comparison of meter verification features and functions: original version vs. Smart Meter 
Verification 
continued
Feature or function
Meter verification application
Original version
Smart Meter Verification