Diagnostics and Troubleshooting
Erratic non-zero flow rate under
no-flow conditions
no-flow conditions
RF interference
Check environment for RF interference.
See Section 22.11.5.
See Section 22.11.5.
Wiring problem
Verify all sensor-to-transmitter wiring
and ensure the wires are making good
contact.
and ensure the wires are making good
contact.
Incorrectly grounded 9-wire cable
(remote core processor with remote
transmitter installations)
(remote core processor with remote
transmitter installations)
Verify 9-wire cable installation.
Vibration in pipeline at rate close to
sensor tube frequency
sensor tube frequency
Check environment and remove source
of vibration.
of vibration.
Improper sensor grounding (T-Series
sensors only)
sensors only)
Verify that the sensor is grounded to
earth ground.
earth ground.
Leaking valve or seal
Check pipeline.
Inappropriate measurement unit
Check configuration. See
Section 22.13.
Section 22.13.
Inappropriate damping value
Check configuration. See Section 7.3.2.
Slug flow
See
Plugged flow tube
Check drive gain and tube frequency.
Purge the flow tubes or replace the
sensor.
Purge the flow tubes or replace the
sensor.
Moisture in sensor junction box
Open junction box and allow it to dry.
Do not use contact cleaner. When
closing, ensure integrity of gaskets and
O-rings, and grease all O-rings.
Do not use contact cleaner. When
closing, ensure integrity of gaskets and
O-rings, and grease all O-rings.
Mounting stress on sensor
Check sensor mounting. Ensure:
• Sensor is not being used to support
pipe.
• Sensor is not being used to correct
pipe misalignment.
• Sensor is not too heavy for pipe.
Sensor cross-talk
Check environment for sensor with
similar (± 0.5 Hz) tube frequency.
similar (± 0.5 Hz) tube frequency.
Incorrect sensor orientation
Sensor orientation must be appropriate
to process fluid. See the installation
manual for your sensor.
to process fluid. See the installation
manual for your sensor.
Erratic non-zero flow rate when flow
is steady
is steady
Output wiring problem
Verify wiring between Series 3000
device and receiving device. See the
transmitter installation instructions.
device and receiving device. See the
transmitter installation instructions.
Problem with receiving device
Test with another receiving device.
Inappropriate measurement unit
Check configuration. See
Section 22.13.
Section 22.13.
Inappropriate damping value
Check configuration. See Section 7.3.2.
Excessive or erratic drive gain
and
.
Slug flow
See
Plugged flow tube
Check drive gain and tube frequency.
Purge the flow tubes or replace the
sensor.
Purge the flow tubes or replace the
sensor.
Wiring problem
Verify all sensor-to-transmitter wiring
and ensure the wires are making good
contact.
and ensure the wires are making good
contact.
Table 22-1
Process variables problems and possible remedies continued
Symptom
Cause
Possible remedy