7.4
Configuring the frequency input
Frequency input parameters define how the frequency input is used to represent process data. 
Frequency input parameters are listed and defined in Table 7-15.
Table 7-14
Discrete input actions and assignments
Action
Default
Assignment
Description
Start zero
None
None
Discrete input 1
Discrete input 2
Discrete event 1
Discrete event 2
Discrete event 3
Discrete event 4
Discrete event 5
Starts the sensor zero process. See Section 16.3.
Reset mass total
None
Resets the value of the mass totalizer to zero.
Reset volume total
None
Resets the value of the liquid volume totalizer to zero.
Reset GStd vol total
None
Resets the value of the gas standard volume totalizer 
to zero.
Reset API corrected 
volume total
(1)
(1) Displayed only if petroleum measurement application is installed.
None
Resets the value of the API corrected volume totalizer 
to zero.
Reset ED standard 
volume total
(2)
(2) Displayed only if enhawnced density application is installed.
None
Resets the value of the ED standard volume totalizer 
to zero.
Reset ED net mass 
total
None
Resets the value of the ED net mass totalizer to zero.
Reset ED net volume 
total
None
Resets the value of the ED net volume totalizer to 
zero.
Reset all totals
None
Resets the value of all totalizers to zero.
Start/stop all totals
None
• If totalizers are stopped, starts all totalizers.
• If totalizers are started, stops all totalizers.
Start meter verify
(3)
(3) Displayed only on systems with Smart Meter Verification.
None
• Starts a Smart Meter Verification test. See 
Table 7-15
Frequency input parameters
Variable
Default
Description
Flow rate units
kg/min
Select the desired unit of mass flow or volume flow. See Table 7-4.
Scaling method
Frequency = flow
• Select frequency = flow, pulses/unit, or units/pulse.
• The frequency input has a range of 0 to 20,000 Hz.
Frequency
1000.000 Hz
If Frequency = flow is selected as the scaling method, enter the frequency (or 
pulse rate), in Hz, that represents the configured flow rate.
Flow
999.9999 kg/min
If Frequency = flow is selected as the scaling method, enter the flow rate that is 
represented by the configured frequency.
Pulses
60.00 pulses
If Pulses/unit is selected as the scaling method, enter the number of input 
pulses that represents one mass or volume unit.
Units
0.0167 kg
If Units/pulse is selected as the scaling method, enter the number of mass or 
volume units that is represented by one input pulse.
K-factor
1.0000
• The K-factor is used for proving a Model 3300 or Model 3350 controller, for 
which meter factors are not available. To calculate the K-factor, see the 
following example. The resulting value must be between 0.0001 and 2.0000.
• The entered value serves as a scaling factor for flow rate outputs and 
displays.