Emerson Model 1700 Manuel D’Utilisation

Page de 350
Options
Description
HI
x > A
The event occurs when the value of the assigned process variable (x) is
greater than the setpoint (Setpoint A), endpoint not included.
LO
x < A
The event occurs when the value of the assigned process variable (x) is
less than the setpoint (Setpoint A), endpoint not included.
IN
A ≤ x ≤ B
The event occurs when the value of the assigned process variable (x) is
“in range,” that is, between Setpoint A and Setpoint B, endpoints included.
OUT
x ≤ A or x ≥ B
The event occurs when the value of the assigned process variable (x) is
“out of range,” that is, less than Setpoint A or greater than Setpoint B, end-
points included.
3.
Assign a process variable to the event.
4.
Set values for the required setpoints.
For HI and LO events, set Setpoint A.
For IN and OUT events, set Setpoint A and Setpoint B.
5.
(Optional) Configure a discrete output to switch states in response to the event
status.
6.
(Optional) Specify the action or actions that the transmitter will perform when the
event occurs.
With the display: OFF-LINE MAINT > OFF-LINE CONFG > IO > CH C > SET DI > DI ACT
With ProLink II: ProLink > Configuration > Discrete Input
With ProLink III: Device Tools > Configuration > I/O > Action Assignment
With the Field Communicator: Configure > Alert Setup > Discrete Events > Assign
Discrete Action
Options for Enhanced Event Action
Options for Enhanced Event Action
Table 6-11:   
Action
Label
Display (stand-
ard)
Chinese-lan-
guage display
ProLink II
ProLink III
Field Communi-
cator
Standard
None (default)
NONE
None
None
None
None
Start sensor zero
START ZERO
Start Sensor Zero
Start Sensor Zero
Start Sensor Zero
Perform auto zero
Start/stop all to-
talizers
START STOP
Start/Stop All Totali-
zation
Start/Stop All Totali-
zation
Start/Stop All Totali-
zation
Start/stop totals
Integrate the meter with the control system
Configuration and Use Manual 
101