Emerson Model 1700 Manuel D’Utilisation

Page de 350
Option
Required values
Sawtooth
Period
Minimum
Maximum
Sine
Period
Minimum
Maximum
4.
For density, set Wave Form as desired and enter the required values.
Option
Required values
Fixed
Fixed Value
Sawtooth
Period
Minimum
Maximum
Sine
Period
Minimum
Maximum
5.
For temperature, set Wave Form as desired and enter the required values.
Option
Required values
Fixed
Fixed Value
Sawtooth
Period
Minimum
Maximum
Sine
Period
Minimum
Maximum
6.
Observe the system response to the simulated values and make any appropriate
changes to the transmitter configuration or to the system.
7.
Modify the simulated values and repeat.
8.
When you have finished testing or tuning, disable sensor simulation.
7.1.1
Sensor simulation
Sensor simulation allows you to test the system or tune the loop without having to create
the test conditions in your process. When sensor simulation is enabled, the transmitter
reports the simulated values for mass flow, density, and temperature, and takes all
appropriate actions. For example, the transmitter might apply a cutoff, activate an event,
or post an alarm.
Completing the configuration
Configuration and Use Manual 
113