Emerson Model 1700 Manuel D’Utilisation

Page de 350
2.4
Characterize the flowmeter (if required)
Display (standard)
Not available
Chinese-language
display
Offline Maintain
 > Configuration > Calibrate Sensor
ProLink II
ProLink
 > Configuration > Device > Sensor Type
ProLink
 > Configuration > Flow
ProLink
 > Configuration > Density
ProLink
 > Configuration > T Series
ProLink III
Device Tools
 > Calibration Data
Field Communicator Configure > Manual Setup > Characterize
Overview
Characterizing the flowmeter adjusts your transmitter to match the unique traits of the
sensor it is paired with. The characterization parameters (also called calibration
parameters) describe the sensor’s sensitivity to flow, density, and temperature.
Depending on your sensor type, different parameters are required. Values for your sensor
are provided by Micro Motion on the sensor tag or the calibration certificate.
Tip
If your flowmeter was ordered as a unit, it has already been characterized at the factory. However,
you should still verify the characterization parameters.
Procedure
1.
Specify Sensor Type.
Straight-tube
Curved-tube
 (all sensors except T-Series)
2.
Set the flow characterization parameters. Be sure to include all decimal points.
For straight-tube sensors, set FCF (Flow Cal or Flow Calibration Factor), FTG, and
FFQ
For curved-tube sensors, set Flow Cal (Flow Calibration Factor).
3.
Set the density characterization parameters.
For straight-tube sensors, set D1, D2, DT, DTG, K1, K2, FD, DFQ1, and DFQ2
For curved-tube sensors, set D1, D2, TC, K1, K2, and FD. (TC is sometimes shown
as DT.)
(5) Devices with the Chinese-language display do not support T-Series sensors.
Quick start
8
 Micro Motion
®
 Model 1700 Transmitters with Analog Outputs