Emerson Model 1700 Manuel D’Utilisation

Page de 350
Procedure
1.
Test the mA output(s).
a. Choose Device Tools > Diagnostics > Testing > mA Output 1 Test or Device Tools >
Diagnostics
 > Testing > mA Output 2 Test.
b. Enter 4 in Fix to:.
c. Click Fix mA.
d. Read the mA current at the receiving device and compare it to the transmitter
output.
The readings do not need to match exactly. If the values are slightly different,
you can correct the discrepancy by trimming the output.
e. Click UnFix mA.
f. Enter 20 in Fix to:.
g. Click Fix mA.
h. Read the mA current at the receiving device and compare it to the transmitter
output.
The readings do not need to match exactly. If the values are slightly different,
you can correct the discrepancy by trimming the output.
i. Click UnFix mA.
2.
Test the frequency output(s).
a. Choose Device Tools > Diagnostics > Testing > Frequency Output Test.
b. Enter the frequency output value in Fix to.
c. Click Fix FO.
d. Read the frequency signal at the receiving device and compare it to the
transmitter output.
e. Click UnFix FO.
3.
Test the discrete output(s).
a. Choose Device Tools > Diagnostics > Testing > Discrete Output Test.
b. Set Fix To: to ON.
c. Verify the signal at the receiving device.
d. Set Fix To: to OFF.
e. Verify the signal at the receiving device.
f. Click UnFix.
Postrequisites
If the mA output reading at the receiving device was slightly inaccurate, you can
correct this discrepancy by trimming the output.
Troubleshooting
Configuration and Use Manual 