Define a special measurement unit for mass flow
Display (standard)
Not available
Chinese-language
display
Offline Maintain
 > Configuration > Units > Special Mass Flow
ProLink II
ProLink
 > Configuration > Special Units
ProLink III
Device Tools
 > Configuration > Process Measurement > Flow > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units > Mass Special Units
Overview
A special measurement unit is a user-defined unit of measure that allows you to report
process data, totalizer data, and inventory data in a unit that is not available in the
transmitter. A special measurement unit is calculated from an existing measurement unit
using a conversion factor.
Note
Although you cannot define a special measurement unit using the display (standard option), you can
use the standard display to select an existing special measurement unit, and to view process data
using the special measurement unit.
Procedure
1.
Specify Base Mass Unit.
Base Mass Unit
 is the existing mass unit that the special unit will be based on.
2.
Specify Base Time Unit.
Base Time Unit
 is the existing time unit that the special unit will be based on.
3.
Calculate Mass Flow Conversion Factor as follows:
a. x base units = y special units
b. Mass Flow Conversion Factor = x/y
4.
Enter Mass Flow Conversion Factor.
5.
Set Mass Flow Label to the name you want to use for the mass flow unit.
6.
Set Mass Total Label to the name you want to use for the mass total and mass
inventory unit.
The special measurement unit is stored in the transmitter. You can configure the
transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for mass flow
You want to measure mass flow in ounces per second (oz/sec).
1.
Set Base Mass Unit to Pounds (lb).
2.
Set Base Time Unit to Seconds (sec).
Configure process measurement
Configuration and Use Manual 
23