Effect of Flow Direction on the mA output: Lower Range Value < 0
Figure 4-2:   
Flow Direction = Forward
mA output
-x
0
x
Reverse flow
Forward flow
20
12
4
Flow Direction = Reverse, Negate Forward
mA output
-x
0
x
Reverse flow
Forward flow
20
12
4
Flow Direction = Absolute Value, Bidirectional, 
Negate Bidirectional
mA output
-x
0
x
Reverse flow
Forward flow
20
12
4
Lower Range Value
 = 
−x
Upper Range Value
 = x
Example: Flow Direction = Forward and Lower Range Value = 0
Configuration:
Flow Direction
 = Forward
Lower Range Value
 = 0 g/sec
Upper Range Value
 = 100 g/sec
Result:
Under conditions of reverse flow or zero flow, the mA output is 4 mA.
Under conditions of forward flow, up to a flow rate of 100 g/sec, the mA output
varies between 4 mA and 20 mA in proportion to the flow rate.
Under conditions of forward flow, if the flow rate equals or exceeds 100 g/sec, the
mA output will be proportional to the flow rate up to 20.5 mA, and will be level at
20.5 mA at higher flow rates.
Example: Flow Direction = Forward and Lower Range Value < 0
Configuration:
Flow Direction
 = Forward
Lower Range Value
 = 
−100 g/sec
Upper Range Value
 = +100 g/sec
Result:
Under conditions of zero flow, the mA output is 12 mA.
Under conditions of forward flow, for flow rates between 0 and +100 g/sec, the mA
output varies between 12 mA and 20 mA in proportion to (the absolute value of) the
flow rate.
Configure process measurement
Configuration and Use Manual 
41