MODEL 5081-P pH/ORP
SECTION 10.0
PROGRAMMING  FOR ORP MEASUREMENTS
10.4 TEMPERATURE ELEMENT
10.4.1 Purpose
This section describes how to match the transmitter to the type of temperature
element in the ORP sensor.
10.4.2 Definition
TEMPERATuRE  ELEMENT: ORP  sensors  use  a  variety  of  temperature  ele-
ments. The Model 5081-P ORP transmitter recognizes the following temperature
elements and configurations:
a. three and four wire 100 ohm platinum RTds
b. three and four wire 1000 ohm platinum RTds
A 100 ohm platinum RTd has a resistance of 100 ohms at 0°C. A 1000 ohm plat-
inum RTd has a resistance of 1000 ohms at 0°C. Although only two lead wires
are  necessary  to  connect  the  RTd  to  the  transmitter,  connecting  a  third  wire
allows  the  transmitter  to  correct  for  the  resistance  of  the  lead  wires  and  for
changes in wire resistance with temperature.
The Model 5081-P transmitter can also be used with a two-wire RTd. Select a
three-wire  configuration  and  jumper  the  RTd  return  and  -RTd  sense  terminals
(terminals 3 and 4, respectively).
10.4.2 Procedure
1. Press PROG on the infrared remote controller (IRC).
2. Press NEXT until the tEMP sub-menu appears in the display. Press
ENTER .
3. Press EXIT to return to the process display.
PROGRAM
tEMP
EXIT
NEXT       ENTER