MODEL 5081-P pH/ORP
SECTION 11.0
MAINTENANCE
TABLE 11-1.  Replacement Parts for Model 5081-P pH/ORP Transmitter
PROBLEM
CLEANING SUGGESTIONS
Loose scale or debris
use a stream of water from a wash bottle to rinse away solids from the tip of the sensor. 
If water does not work, gently wipe the glass bulb and liquid junction with a soft cloth, 
tissue, cotton-tipped swab, or a soft bristle brush.
Oil and grease
Wash the glass bulb with mild detergent solution and rinse thoroughly with water.
Hard scale (carbonate
If wiping the sensor tip with a tissue or cotton swab does not remove the scale, soak the
sulfate scales and 
glass bulb ONLY in a solution of 5% hydrochloric acid. To prepare the acid solution, add
corrosion products) 
15 mL of concentrated hydrochloric acid to 85 mL of water. Keep the acid away from the 
liquid junction and from any stainless steel portions of the sensor. Rinse the  sensor 
thoroughly with deionized water. Some scales (for example, calcium sulfate) cannot be 
removed easily with acid. Soaking the glass bulb in a 2% solution of disodium EdTA may 
be helpful.
PN
Description
Shipping 
weight
23992-02
For Model 5081-P-HT — PCB stack consisting of the CPu, communication, and analog
1 lb/0.5 kg
boards; display board is not included; CPu, communication, and analog boards are
factory-calibrated as a unit and cannot be ordered separately
23992-03
For Model 5081-P-FF — PCB stack consisting of the CPu, communication, and analog
1 lb/0.5 kg
boards; display board is not included; CPu, communication, and analog boards are
factory-calibrated as a unit and cannot be ordered separately
23652-01
LCd display PCB
1 lb/0.5 kg
33337-02
Terminal block
1 lb/0.5 kg
23593-01
Enclosure cover, front with glass window
3 lb/1.5 kg
33360-00
Enclosure, center housing
4 lb/1.5 kg
33362-00
Enclosure cover, rear
3 lb/1.0 kg
9550187
O-ring (2-252), one, front and rear covers each require an O-ring
1 lb/0.5 kg
NOTE
Screw, 8-32 x 0.5 inch, for attaching terminal block to center housing
*
NOTE
Screw, 8-32 x 1.75 inch, for attaching circuit board stack to center 
*
housing
33342-00
Cover lock
1 lb/0.5 kg
33343-00
Locking bracket nut
1 lb/0.5 kg
NOTE
Screw, 10-24 x 0.38 inch, for attaching cover lock and locking bracket nut to center housing
*
NOTE:  For information only. Screws cannot be purchased from Rosemount Analytical.
* Weights are rounded up to the nearest whole pound or 0.5 kg.
11.3  pH SENSOR MAINTENANCE
11.3.1  Frequency of Cleaning
The frequency at which a sensor should be inspected and cleaned can be determined only by experience. If the process
liquid coats or fouls the sensor, frequent cleaning may be necessary. If the process does not contain a high level of sus-
pended solids, the need for regular cleaning will be less. Often an increase in glass impedance indicates the electrode is
becoming fouled and needs cleaning. Refer to Section 12.4 for a description of the glass impedance diagnostic.
11.3.2  Cleaning Procedures