MODEL 5081-P pH/ORP
SECTION 12.0
TROUBLESHOOTING
12.2 CALIBRATION ERRORS 
If an error occurs during calibration, an error message appears in the main display, and the transmitter does not update
the calibration. The calibration errors are Std ErrSLOPE Err LO, and SLOPE Err HI. See Section 12.4 for an explana-
tion of the error messages and suggested ways of correcting the problem.
12.3  TROUBLESHOOTING - GENERAL
Troubleshooting is easy as 1, 2, 3 . . .
Step 1 Look for a diagnostic message on the  display to help identify the problem. Refer to Section 12.4 for an explana-
tion of the message and a list of the possible problems that triggered it.
Step 2 Refer to Section 12.5 for common measurement problems and the  recommended actions to resolve them.
Step 3 Follow the step by step troubleshooting approach, offered in Section 12.6, to diagnose and correct less common
or more complex  problems.
12.4  TROUBLESHOOTING wHEN A DIAGNOSTIC MESSAGE IS SHOwING
The Model 5081-P pH/ORP transmitter continuously monitors the measurement loop (sensor and transmitter) for prob-
lems. If a problem is detected, the transmitter displays a fault or error message. The message appears in the tempera-
ture/output area of the main display. The table lists each diagnostic message and the section to consult for help.
MESSAGE
SECTION
GLASSFAIL
12.4.1
GLASSWArn
12.4.2
rEF FAIL
12.4.3
rEF WArn
12.4.4
CALIbrAtE
12.4.5
tEMP HI
12.4.6
tEMP LO
12.4.6
LInE FAIL
12.4.7
InPut WArn
12.4.8
SLOPE Err LO
12.4.9
SLOPE Err HI
12.4.10
Std Err
12.4.11
rOM FAIL
12.4.12
CPu FAIL
12.4.12
AdC WArn
12.4.13
CyCLE PWr
12.4.13
WrItE Err
12.4.14
FACt FAIL
12.4.15