Emerson 5081-P Manuel D’Utilisation

Page de 156
122
MODEL 5081-P pH/ORP
SECTION 12.0
TROUBLESHOOTING
SyMPTOM
SECTION
Id 000 appears in display when trying to program or calibrate transmitter
12.5.1
Error message flashing in display
12.4
Transmitter does not respond to remote controller
12.5.2
Calibration Problems:
SLOPE Err HI or SLOPE Err LO appears after calibration attempt
12.5.3
bF1 or bF2 continuously flashes during auto calibration
12.5.4
pH reading in buffer drifts during manual calibration
12.5.5
Measurement Problems:
Sensor does not respond to known pH changes
12.5.6
Buffer calibration is acceptable; process pH is slightly different from expected value
12.5.7
Buffer calibration is acceptable; process pH is grossly wrong and/or readings are noisy
12.5.8
Temperature reading is inaccurate
12.5.9
Transmitter problems
No display
12.5.10
display segments missing or display incorrect
12.5.11
Transmitter locked up, all display segments lit
12.5.12
Transmitter periodically restarts itself
12.5.13
12.4.15 FACt FAIL
FACt FAIL appears if the transmitter factory calibration message has been triggered. A stray noise spike can cause this
message to appear. If the pH reading seems acceptable, reset the calibration flag.
1. Enter the factory calibration menu by pressing  ç on the IRC ten times. The display will not change. Immediately 
press é. FActoryCAL appears in the display.
2. Press  NEXT.  rEPAir appears in the display.
3. Press  NEXT.  ConFiG appears in the display.
4. Press  NEXT.  rESEt appears in the display.
5. Press  ENTER.  rESEtCFG appears in the display.
6. Press  ENTER.  rESEt appears again.
7. Press  NEXT.  FActoryCAL reappears.
8. Press  ENTER.  FactOn appears in the display.
9. Press  é.  FactOFF appears. Press  ENTER  to store the settings.
10. Press  EXIT repeatedly until the main display reappears.
If the message does not clear or problems persist, the electronics have failed. Replace the electronic board stack.
12.5  TROUBLESHOOTING wHEN NO DIAGNOSTIC MESSAGE IS SHOwING
If no diagnostic message is showing, locate the symptom(s) in the table below and refer to the appropriate section for
assistance.