Emerson 5081-P Manuel D’Utilisation

Page de 156
134
MODEL 5081-P pH/ORP
SECTION 13.0 
pH MEASUREMENTS
The second term, 0.1984 T pH, is the potential (in mV) at
the outside surface of the pH glass. This potential depends
on  temperature  and  on  the  pH  of  the  sample. Assuming
temperature remains constant, any change in cell voltage
is  caused  solely  by  a  change  in  the  pH  of  the  sample.
Therefore, the cell voltage is a measure of the sample pH.
Note that a graph of equation 1, E(T) plotted against pH,
is a straight line having a y-intercept of E°(T) and a slope
of 0.1984 T. 
13.6  GLASS ELECTRODE SLOPE
For reasons beyond the scope of this discussion, equation
1  is  commonly  rewritten  to  remove  the  temperature
dependence  in  the  intercept  and  to  shift  the  origin  of  the
axes to pH 7. The result is plotted in Figure 13-6. Two lines
appear  on  the  graph.  One  line  shows  how  cell  voltage
changes with pH at 25°C, and the other line shows the rela-
tionship  at  50°C.  The  lines,  which  are  commonly  called
isotherms,  intersect  at  the  point  (pH  7,  0  mV). An  entire
family  of  curves,  each  having  a  slope  determined  by  the
temperature and all passing through the point (pH 7,  0 mV)
can be drawn on the graph.
Figure 13-6 shows why temperature is important in making
pH  measurements.  When  temperature  changes, the  slope
of  the  isotherm  changes.  Therefore,  a  given  cell  voltage
corresponds to a different pH value, depending on the tem-
perature. For example, assume the cell voltage is -150 mV.
At  25°C  the  pH  is  9.54,  and  at  50°C  the  pH  is  9.35. The
process  of  selecting  the  correct  isotherm  for  converting
voltage to pH is called temperature compensation. All mod-
ern  process  pH  meters,  including  the  Model  5081-P
pH/ORP transmitter, have automatic temperature compen-
sation.
The slope of the isotherm is often called the glass electrode
or  sensor  slope.  The  slope  can  be  calculated  from  the
equation: slope = 0.1984 (t + 273.15), where t is tempera-
ture in °C. The slope has units of mV per unit change in pH.
The table lists slopes for different temperatures.
Temp (°C)
Slope (mv/unit pH)
15
-57.2
20
-58.2
25
-59.2
30
-60.1
35
-61.1
As the graph in Figure 13-6 suggests, the closer the pH is
to 7, the less important is temperature compensation. For
example, if the pH is 8 and the temperature is 30°C, a 10°C
error in temperature introduces a pH error of ±0.03. At pH
10, the error in the measured pH is ±0.10. 
12.7  BUFFERS AND CALIBRATION
Figure 13-6 shows an ideal cell: one in which the voltage is
zero when the pH is 7, and the slope is 0.1984 T over the
entire pH range. In a real cell the voltage at pH 7 is rarely
zero,  but  it  is  usually  between  -30  mV  and  +30  mV.  The
slope is also seldom 0.1984 T over the entire range of pH.
However, over a range of two or three pH units, the slope
is usually close to ideal.
Calibration compensates for non-ideal behavior. Calibration
involves the use of solutions having exactly know pH, called
calibration buffers or simply buffers. Assigning a pH value to
a  buffer  is  not  a  simple  process.    The  laboratory  work  is
demanding,  and  extensive  theoretical  work  is  needed  to
support certain assumptions that must be made. Normally,
establishing  pH  scales  is  a  task  best  left  to  national  stan-
dards  laboratories.  pH  scales  developed  by  the  united
States  National  Institute  of  Standards  and  Technology
(NIST),  the  British  Standards  Institute  (BSI),  the  Japan
Standards  Institute  (JSI),  and  the  German  deutsche
Institute  für  Normung  (dIN)  are  in  common  use. Although
there are some minor differences, for practical purposes the
scales are identical. Commercial buffers are usually trace-
able to a recognized standard scale. Generally, commercial
buffers  are  less  accurate  than  standard  buffers.  Typical
accuracy is ±0.01 pH units. Commercial buffers, sometimes
called  technical  buffers,  do  have  greater  buffer  capacity.
They are less susceptible to accidental contamination and
dilution than standard buffers. 
Figure  13-7  shows  graphically  what  happens  during  cali-
bration. The example assumes calibration is being done at
pH 7.00 and pH 10.00. When the electrodes are placed in
pH 7 buffer the cell voltage is V
7
, and when the electrodes
FIGURE 13-6. Glass Electrode Slope.
The voltage of a pH measurement cell depends on pH and
temperature.  A  given  pH  produces  different  voltages
depending on the temperature. The further from pH 7, the
greater  the  influence  of  temperature  on  the  relationship
between pH and cell voltage.