15
MODEL 5081-P pH/ORP
SECTION 2.0
INSTALLATION
2.6 POwER SUPPLy wIRING FOR 
MODEL 5081-P-FF
2.6.1  Power  Supply  wiring.    Refer  to  Figure  2-5  and
Figure 2-6.
Run the power/signal wiring through the opening nearest
terminals 15 and 16. use shielded cable and ground the
shield  at  the  power  supply.  To  ground  the  transmitter,
attach the shield to the grounding screw on the inside of
the transmitter case. A third wire can also be used to con-
nect the transmitter case to earth ground.
NOTE
For optimum EMI/RFI immunity, the power sup-
ply/output  cable  should  be  shielded  and
enclosed in an earth-grounded metal conduit.
do not run power supply/signal wiring in the same con-
duit or cable tray with AC power lines or with relay actu-
ated  signal  cables.  Keep  power  supply/signal  wiring  at
least 6 ft (2 m) away from heavy electrical equipment. 
FIGURE 2-5. Typical Fieldbus Network Electrical
wiring Configuration
FIGURE 2-6. Model 5081-P-FF Power wiring Details
5081-P pH/ORP
Transmitter
5081-P pH/ORP
Transmitter