MODEL 5081-P pH/ORP
SECTION 5.0 
OPERATION wITH REMOTE CONTROLLER
FIGURE 5-3. Program Display Screen
The program display screen appears when calibrating, programming, or reading diagnostic messages. 
FIGURE 5-2. Process Display Screen
The process display screen appears during normal operation.
pH or ORP (ORP in mV)
Temperature in °C or °F
5.2 DISPLAyS
Figure 5-2 shows the process display screen, and Figure 5-3 shows the program display screen.
5.1.1 Software Functionality. The Model 5081-P pH/ORP software is designed to permit remote testing and
configuration of the transmitter using the Fisher-Rosemount deltaV Fieldbus Configuration Tool, or other 
F
OuN-
dATION fieldbus compliant host.
5.1.2 Transducer Block.  The transducer block contains the actual measurement data. It includes information
about sensor type, engineering units, reranging, damping, temperature compensation, calibration, and diagnostics.
5.1.3 Resource Block.  The resource Block contains physical device information, including available memory,
manufacturer identification, type of device, and features.
5.1.4  
F
OUNDATION fieldbus Function Blocks.  The Model 5081-P pH/ORP includes three Analog Input (AI)
function blocks and one Input Selector (ISEL) function block as part of its standard offering.
Analog Input. The Analog Input (AI) block processes the measurement and makes it available to other func-
tion blocks. It also allows filtering, alarming, and engineering unit change.
Charaterizer (optional). The characterizer block changes the characteristic of the input signal. Common
uses of the characterizer block include converting temperature to density or humidity, and converting milli-
volts to temperature for an IR sensor.