Emerson 00809-0100-4514 Manuel D’Utilisation

Page de 60
 
IMPORTANT NOTICE -- ERRATA 
Model 1154 Product Manual 00809-0100-4514 Rev BA (January 2008) 
No. 
Affected 
Pages 
Description of Change 
Effect. 
Date 
1 6-6 
Process Flange –CF3M (Cast version of 316L SST) 
Drain/Vent Valves –316L SST 
Process Connections – 3/8-inch Swagelok compression fitting, 316L SST (1/4-18 
NPT optional) 
10/21/09 
3-5 
Change the last paragraph on the page to read as follows: 
”Damping electronics are available as an option.  Transmitters with standard 
electronics can be retrofitted with the adjustable damping feature by changing out 
both the amplifier board and the calibration board.  Please reference Table 6-2 
(Rosemount 1154DP, 1154HP, and 1154GP Spare Parts) for the applicable part 
numbers.” 
4/13/12 
6-9 
Table 6-2,”Rosemount 1154DP, 1154HP, and 1154GP Spare Parts”, in all locations 
the following part numbers are updated: 
Amplifier Circuit Board, Output Code R:  
”01154-0001-0005” is replaced by ”01154-0153-0001” 
Amplifier Circuit Board with Damping, Output Code R: 
”01154-0021-0004” is replaced by ”01154-0156-0001” 
Amplifier Circuit Board, Output Code R, N0026: 
”01154-0001-0006” is replace by ”01154-0153-0002” 
Sensor Module, 316 SST 
(5)
0-25/150 inH
2
O:  
”01154-0300-0242” is replaced by ”01154-5300-0242” 
 
 
”01154-0300-0342” is replaced by ”01154-5300-0342” 
  
 
”01154-0300-0142” is replaced by ”01154-5300-0142” 
0-125/750 inH
2
O:  ”01154-0300-0252” is replaced by ”01154-5300-0252” 
 
 
”01154-0300-0352” is replaced by ”01154-5300-0352” 
 
 
”01154-0300-0152” is replaced by ”01154-5300-0152” 
0-17/100 psi: 
”01154-0300-0262” is replaced by ”01154-5300-0262” 
 
”01154-0300-0362” is replaced by ”01154-5300-0362” 
 
”01154-0300-0162” is replaced by ”01154-5300-0162” 
0-50/300 psi: 
”01154-0300-0272” is replaced by ”01154-5300-0272” 
 
”01154-0300-0372” is replaced by ”01154-5300-0372” 
 
”01154-0300-0172” is replaced by ”01154-5300-0172” 
0-170/1,000 psi: 
”01154-0300-0282” is replaced by ”01154-5300-0282” 
 
”01154-0300-0182” is replaced by ”01154-5300-0182” 
0-500/3,000 psi: 
”01154-0300-0192” is replaced by ”01154-5300-0192” 
0-1,000/4,000 psi:  ”01154-0300-0102” is replaced by ”01154-5300-0102” 
 
4/13/12 
6-10 
Table 6-2,” Rosemount 1154DP, 1154HP, and 1154GP Spare Parts”, the table 
footnotes are updated to add note (5) which will read as follows: 
(5) IMPORTANT NOTICE: To maintain a transmitter’s qualified configuration, 
when purchasing or installing a new Sensor Module, Rosemount Qualification 
report D2011019 must be carefully reviewed to verify that the Sensor Module 
to be installed and the associated Amplifier Circuit Board in a given 
transmitter is a qualified configuration
.  As detailed in the referenced report, not 
all Sensor Module part numbers are considered qualified in conjunction with certain 
Amplifier Circuit Boards. 
4/13/12