Emerson 9739 Manuel D’Utilisation

Page de 256
Important
Whenever you change a frequency output parameter, verify all other frequency output parameters
before returning the flowmeter to service. In some situations, the transmitter automatically loads a
set of stored values, and these values may not be appropriate for your application.
6.2.1
Configure Frequency Output Power Source
Display
OFF-LINE MAINT
 > OFF-LINE CONFG > IO > FO > POWER
ProLink II
ProLink
 > Configuration > Frequency > Power Type
Field Communicator Configure > Manual Setup > Inputs/Outputs > Frequency Output > FO Settings > Power Source
Overview
Use Frequency Output Power Source to set the power source for the frequency output. The
power configuration must match the wiring for the frequency output.
Procedure
Set Frequency Output Power Source as desired.
Option
Description
Internal
The output is powered by the transmitter.
External
The output is powered by an external power source.
6.2.2
Configure Frequency Output Process Variable
Display
OFF-LINE MAINT
 > OFF-LINE CONFG > IO > FO
ProLink II
ProLink
 > Configuration > Frequency
ProLink III
Device Tools
 > Configuration > I/O > Outputs > Frequency Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > Frequency Output
Overview
Frequency Output Process Variable
 controls the variable that is reported over the frequency
output.
Prerequisites
If you plan to configure the output to report volume flow, ensure that you have set Volume
Flow Type
 as desired: Liquid or Gas Standard Volume.
If you plan to configure an output to report a concentration measurement process
variable, ensure that the concentration measurement application is configured so that the
desired variable is available.
Integrate the meter with the control system
92
Micro Motion
®
 9739 MVD Transmitters