Emerson 9739 Manuel D’Utilisation

Page de 256
8.6
Reset totalizers
Display
See 
.
ProLink II
ProLink
 > Totalizer Control > Reset Mass Total
ProLink
 > Totalizer Control > Reset Volume Total
ProLink
 > Totalizer Control > Reset Gas Volume Total
ProLink
 > Totalizer Control > Reset
Field Communicator Service Tools > Variables > Totalizer Control > Mass > Mass Total
Service Tools
 > Variables > Totalizer Control > Gas Standard Volume > Volume Total
Service Tools
 > Variables > Totalizer Control > Gas Standard Volume > GSV Total
Service Tools
 > Variables > Totalizer Control > All Totalizers > Reset All Totals
Overview
When you reset a totalizer, the transmitter sets its value to 0. It does not matter whether
the totalizer is started or stopped. If the totalizer is started, it continues to track process
measurement.
Tip
When you reset a single totalizer, the values of other totalizers are not reset. Inventory values are not
reset.
8.6.1
Reset totalizers using the display
Prerequisites
The Totalizer Reset display function must be enabled.
The totalizer that you want to reset must be configured as a display variable. For example:
If you want to reset the mass totalizer, Mass Total must be configured as a display
variable.
If you want to reset the volume totalizer, Volume Total must be configured as a display
variable.
Procedure
To reset the mass totalizer:
1. Scroll until the mass totalizer value appears.
2. Select.
3. Scroll until RESET appears beneath the current totalizer value.
4. Select.
5. Select again to confirm.
Transmitter operation
Configuration and Use Manual 