Emerson 9739 Manuel D’Utilisation

Page de 256
2.
Set the flow characterization parameters. Be sure to include all decimal points.
For straight-tube sensors, set FCF (Flow Cal or Flow Calibration Factor), FTG, and FFQ.
For curved-tube sensors, set Flow Cal (Flow Calibration Factor).
3.
Set the density characterization parameters.
For straight-tube sensors, set D1, D2, DT, DTG, K1, K2, FD, DFQ1, and DFQ2.
For curved-tube sensors, set D1, D2, TC, K1, K2, and FD. (TC is sometimes shown
as DT.)
2.4.1
Sample sensor tags
Tag on older curved-tube sensors (all sensors except T-Series)
Figure 2-1:   
Tag on newer curved-tube sensors (all sensors except T-Series)
Figure 2-2:   
Quick start
8
Micro Motion
®
 9739 MVD Transmitters