If LD Optimization is enabled on your meter, disable it. To do this, choose Configure >
Manual Setup
Manual Setup
> Measurements > LD Optimization. LD Optimization is used only with large
sensors in hydrocarbon applications. In some installations, only Micro Motion
customer service has access to this parameter. If this is the case, contact
Micro Motion before continuing.
customer service has access to this parameter. If this is the case, contact
Micro Motion before continuing.
•
The calibrations must be performed without interruption, in the order shown. Make
sure that you are prepared to complete the process without interruption.
sure that you are prepared to complete the process without interruption.
•
Before performing the calibration, record your current calibration parameters. If the
calibration fails, restore the known values.
calibration fails, restore the known values.
Restriction
For T-Series sensors, the D1 calibration must be performed on air and the D2 calibration must be
performed on water.
performed on water.
Procedure
Measurement support
152