Emerson 9739 Manuel D’Utilisation

Page de 256
Status LED states
Table 10-1:   
LED behavior
Alarm condition
Description
Solid green
No alarm
Normal operation
Flashing yellow
No alarm
Zero calibration procedure is in progress
Solid yellow
Low-severity alarm
Alarm condition that will not cause measure-
ment error (outputs continue to report proc-
ess data)
Solid red
High-severity alarm
Alarm condition that will cause measurement
error (outputs in fault)
10.2
Status alarms
Status alarms and recommended actions
Table 10-2:   
Alarm code
Description
Cause
Recommended actions
A001
EEPROM Error (Core
Processor)
An uncorrectable
checksum mismatch
has been detected.
Cycle power to the meter.
Contact Micro Motion.
A002
RAM Error (Core Pro-
cessor)
ROM checksum error or
a RAM location cannot
be written to.
Cycle power to the meter.
Contact Micro Motion.
A003
No Sensor Response
Continuity failure of
drive circuit, LPO, or
RPO, or LPO-RPO mis-
match when driving.
Check the drive gain and the pickoff voltage.
See 
Check the wiring between the sensor and
transmitter. See 
.
.
Check the integrity of the sensor tubes.
A004
Temperature Over-
range
Combination of A016
and A017.
Check the sensor wiring. See 
.
Check the wiring between the sensor and
transmitter. See 
.
Verify temperature characterization parame-
ters (Temp Cal Factor).
Check your process conditions against the val-
ues reported by the flowmeter.
Contact Micro Motion.
Troubleshooting
160
Micro Motion
®