Flow measurement problems and recommended actions
 (continued)
Table 10-3:   
Problem
Possible causes
Recommended actions
Inaccurate flow rate
or batch total
Wiring problem
Inappropriate measurement unit
Incorrect flow calibration factor
Incorrect meter factor
Incorrect density calibration factors
Incorrect flowmeter grounding
Slug flow
Problem with receiving device
Check the wiring between the sensor and
transmitter. See 
.
Verify that the measurement units are con-
figured correctly for your application.
Verify that all of the characterization parame-
ters match the data on the sensor tag.
Zero the meter.
.
Check for slug flow. See 
Verify that the receiving device, and the wir-
ing between the transmitter and the receiv-
ing device.
Check sensor coil resistance and for shorts to
case. See 
Replace the core processor or transmitter.
10.4
Density measurement problems
Density measurement problems and recommended actions
Table 10-4:   
Problem
Possible causes
Recommended actions
Inaccurate density
reading
Problem with process fluid
Incorrect density calibration factors
Wiring problem
Incorrect flowmeter grounding
Slug flow
Plugged or coated flow tube
Incorrect sensor orientation
RTD failure
Physical characteristics of sensor have
changed
Check the wiring between the sensor and
transmitter. See 
.
.
Check your process conditions against the
values reported by the flowmeter.
Verify that all of the characterization parame-
ters match the data on the sensor tag.
Check for slug flow. See 
If two sensors with similar frequency are too
near each other, separate them.
Purge the flow tubes.
Unusually high densi-
ty reading
Plugged or coated flow tube
Incorrect K2 value
Incorrect temperature measurement
RTD problem
In high frequency meters, this can be an
indication of erosion or corrosion
In low frequency meters this can indi-
cate tube fouling
Verify that all of the characterization parame-
ters match the data on the sensor tag.
Purge the flow tubes.
Check for coating in the flow tubes.