Emerson 9739 Manuel D’Utilisation

Page de 256
Transmitter default values and ranges
 (continued)
Table D-1:   
Type
Parameter
Default
Range
Comments
Discrete input
Actions
None
Polarity
Active low
mA input
Process Variable (PV)
None
Display
Backlight on/off
On
Backlight intensity
63
0 – 63
Refresh rate
200 millisec-
onds
100 – 10,000
milliseconds
Variable 1
Mass flow rate
Variable 2
Mass total
Variable 3
Volume flow
rate
Variable 4
Volume total
Variable 5
Density
Variable 6
Temperature
Variable 7
Drive gain
Variable 8–15
None
Display totalizer start/stop
Disabled
Display totalizer reset
Disabled
Display auto scroll
Disabled
Display offline menu
Enabled
Display offline password
Disabled
Display alarm menu
Enabled
Display acknowledge all alarms
Enabled
Offline password
1234
Auto scroll rate
10 sec
Digital com-
munications
Fault action
None
Fault timeout
0 seconds
0.0 – 60.0 sec
Modbus address
1
Modbus ASCII support
Enabled
Floating-point byte order
3–4–1–2
Default values and ranges
Configuration and Use Manual 
233