Status alarms and Status Alarm Severity
 (continued)
Table 5-2:   
Alarm code
Status message
Default severity Notes
Configurable?
A011
Zero Calibration Failed:
Low
Fault
Yes
A012
Zero Calibration Failed:
High
Fault
Yes
A013
Zero Calibration Failed:
Unstable
Fault
Yes
A014
Transmitter Failure
Fault
No
A016
Sensor RTD Failure
Fault
Yes
A017
T-Series RTD Failure
Fault
Yes
A020
No Flow Cal Value
Fault
Yes
A021
Incorrect Sensor Type (K1) Fault
No
A102
Drive Overrange
Informational
Yes
A104
Calibration in Progress
Informational
Can be set to either Informational or
Ignore
, but cannot be set to Fault.
Yes
A105
Slug Flow
Informational
Yes
A107
Power Reset Occurred
Informational
Normal transmitter behavior; oc-
curs after every power cycle.
Yes
A113
mA Output 2 Saturated
Informational
Can be set to either Informational or
Ignore
, but cannot be set to Fault.
Yes
A114
mA Output 2 Fixed
Informational
Can be set to either Informational or
Ignore
, but cannot be set to Fault.
Yes
A116
Temperature Overrange
(Petroleum)
Informational
Applies only to transmitters with
the petroleum measurement appli-
cation.
Yes
A117
Density Overrange (Petro-
leum)
Informational
Applies only to transmitters with
the petroleum measurement appli-
cation.
Yes
A120
Curve Fit Failure (Concen-
tration)
Informational
Applies only to transmitters with
the concentration measurement
application.
No
A121
Extrapolation Alarm (Con-
centration)
Informational
Applies only to transmitters with
the concentration measurement
application.
Yes
A132
Sensor Simulation Active
Informational
Applies only to flowmeters with the
enhanced core processor.
Can be set to either Informational or
Ignore
, but cannot be set to Fault.
Yes
Configure device options and preferences
78
Micro Motion
®
 9739 MVD Transmitters