Emerson 1700 Manuel D’Utilisation

Page de 80
36
Micro Motion
®
 Model 1700 and 2700 Transmitters
Output Wiring – Model 1700/2700 Intrinsically Safe Transmitters
Figure 5-4
Safe area HART multidrop wiring with SMART FAMILY
 transmitters and a configuration tool
5.3.2
Safe area frequency/discrete output wiring
The following frequency/discrete output wiring diagram is an example of proper basic wiring for the 
Model 1700 transmitter’s frequency output or the Model 2700 transmitter’s frequency/discrete output.
Figure 5-5
Safe area frequency/discrete output wiring
HART-compatible
host or controller
HART-compatible 
transmitters
SMART FAMILY
 
transmitter
Note: For optimum HART communication, make sure the output loop 
is single-point-grounded to an instrument-grade ground.
24 VDC loop power 
supply required for 
HART 4–20 mA 
passive transmitters
600 
Ω maximum resistance
250 
Ω minimum resistance
Model 1700 or 2700 
I.S. transmitter
+
Note: See Figure 5-6 for voltage and resistance values.
Counter
VDC
R
load
+
+
00042