4.3.3
Selecting the temperature range
A DAU can be configured for one of three different temperature ranges 
by means of a multiplexer. The multiplexer for the RTDs has two different 
amplification factors which are jumper selected. It is also possible to 
provide an offset to the measured signal by adding a current to it. 
Figure 4-4.
Selecting the temperature range.
The amplification is selected on the X1 and X2 sockets and the offset is 
selected on the X3 according to table 11.3. The standard setting is from -
50°C to +125°C. The database setting of the temperature range in the 
TankMaster WinSetup must correspond to the setting of the jumpers. 
Table 4-1.
Temperature range settings.
A high precision reference resistor must be connected to the X21 
terminal to positions 43, 44 and 45. Depending on the temperature range 
stated in the Required System Information form, one out of three 
different reference resistors is chosen and connected. See Figure
 
and the spare parts list in Rosemount TankRadar REX Service Manual.
Temperature Range 
Pt 100
X1
X2
X3
-50 to +125
°C
OPEN
OPEN
CLOSED
-50 to +300
°C
CLOSED
CLOSED
CLOSED
-200 to +150
°C
CLOSED
CLOSED
OPEN
X1 - OPEN
X2 - OPEN
X3 - CLOSED
DAU_T
empRange.eps