Users Manual 
6-14 
ALARm:BEEP <bool> 
Sets the state of the audible alarm enable. 
Example: ALAR:BEEP 1 
The <bool> parameter turns the alarm beep enable on (1 or ON) or off (0 or OFF). This 
command may be password protected (see Password Commands). 
ALARm:CLEar 
Clears the alarm events. 
Example: ALAR:CLE 
The alarm clear command clears the alarm events. 
ALARm:DATE:FIRSt? 
Returns the date of the first alarm event. 
Example: ALAR:DATE:FIRS? 
Response: 2003,8,10 
The response is returned in the format, <year>,<month>,<day>. 
If no alarm has been triggered, the response is: 2000,0,0. 
ALARm:DATE:LAST? 
Returns the date of the last alarm event. 
Example: ALAR:DATE:LAST? 
Response: 2003,8,10 
The response is returned in the format, <year>,<month>,<day>. 
If no alarm has been triggered, the response is: 2000,0,0. 
ALARm:DISPlay? 
Returns the state of the visible alarm enable. 
Example: ALAR:DISP? 
Response: 0 
A value of 1 is returned if the alarm display is enabled. A value of 0 is returned if the 
alarm display is disabled. 
ALARm:DISPlay <bool> 
Sets the state of the visible alarm enable. 
Example: ALAR:DISP 1 
The <bool> parameter turns the alarm display enable on (1 or ON) or off (0 or OFF). 
This command may be password protected (see Password Commands). 
ALARm:POWer? 
Indicates whether a power loss alarm event has occurred. 
Example: ALAR:POW?