Emerson Rosemount 2230 Manuel D’Utilisation

Page de 80
Reference Manual 
300560EN, Rev AA
May 2011
5-3
Rosemount 2230
Section 5. Service and Troubleshooting
5.2.2
Viewing Input 
and Holding 
Registers
Measured data is continuously stored in the Rosemount 2230 Input 
Registers.
 They can be used for verifying that the Rosemount 2230 works 
properly and for advanced troubleshooting.
The Holding Registers store various configuration parameters which are 
used to control the display presentation.
By using the TankMaster WinSetup configuration tool most holding registers 
can be edited by simply typing a new value in the appropriate value input field. 
To view Input and Holding registers for the 2230 display, do the following:
1. Start the TankMaster WinSetup program
2. In the TankMaster WinSetup workspace window, click the right mouse 
button on the ATD device icon. The ATD device represents all the 
non-level devices such as the 2230 display.
3. Choose the View Input Registers (or View Holding Registers) option, or 
from the Service menu choose Devices > View Input / View Holding 
Registers.