Anritsu S251C Manuel D’Utilisation
Range Errors
A listing of Range Error messages is given in Table 2-3.
2 - 17
Chapter 2 Functions and Operations
Error Message
Description
RANGE
ERROR:F1 > F2
ERROR:F1 > F2
The start (F1) frequency is greater than the stop (F2) frequency.
RANGE
ERROR:D1 > D2
ERROR:D1 > D2
The start (D1) distance is greater than the stop (D2) distance.
RANGE
ERROR:D2 >
DMax=xx.x ft (m)
ERROR:D2 >
DMax=xx.x ft (m)
The stop distance (D2) exceeds the maximum unaliased range. This
range is determined by the frequency span, number of points,
and relative propagation velocity:
range is determined by the frequency span, number of points,
and relative propagation velocity:
MaximumUnaliased Range
dp
V f
F
F
=
´
-
-
( .
) (
) (
)
1 5 108
1
2
1
Where:
dp is the number of data points (130, 259, or 517)
V
V
f
is the relative propagation velocity
F2 is the stop frequency in Hz
F1 is the start frequency in Hz
Maximum Unaliased Range is in meters
F1 is the start frequency in Hz
Maximum Unaliased Range is in meters
Refer to Figure -, page , for a plot of maximum range versus fre-
quency span.
quency span.
RANGE ERROR:
TOP<=BOTTOM
TOP<=BOTTOM
The SWR scale parameter top value is less than or equal to its bottom
value.
value.
RANGE ERROR:
TOP>=BOTTOM
TOP>=BOTTOM
The RL scale parameter top value is greater than or equal to its bottom
value.
value.
CAL
INCOMPLETE
INCOMPLETE
A complete open, short, and load calibration must be performed bef-
ore calibration can be turned on.
ore calibration can be turned on.
DIST REQUIRES
F1 < F2
F1 < F2
Valid distance to fault plots require a non-zero frequency span.
DIST REQUIRES
CAL
CAL
Distance-to-fault measurements cannot be performed with CAL OFF.
USE OPTIONS
MENU TO
SELECT A
PRINTER
MENU TO
SELECT A
PRINTER
Attempting to print a display with no printer selected. Select a printer,
then retry.
then retry.
CANNOT ZERO
NO DETECTOR
INSTALLED
NO DETECTOR
INSTALLED
Attempting to perform a Power Monitor zero adjust function with no
RF detector connected to the Site Master.
RF detector connected to the Site Master.
CANNOT ZERO
INPUT SIGNAL
TOO HIGH
INPUT SIGNAL
TOO HIGH
Attempting to perform a Power Monitor zero adjust function with an
input of greater than –20 dBm.
input of greater than –20 dBm.
POWER MONITOR
OPTION NOT
INSTALLED
OPTION NOT
INSTALLED
Attempting to enter Power Monitor mode with no Option 5 installed.
Table 2-3.