Fluke PM6685R Manuel D’Utilisation

Page de 276
410
‘ —— CHECK MESSAGE AVAILABLE BIT ——
420
WHILE SPR% AND 16
430
PRINT “ Message available bit set”
440
MSG$ = SPACE$(255)
450
CALL IBRD (CNT%, MSG$)
460
LFPOS = INSTR(MSG$, CHR$(10))
470
IF LFPOS <> 0 THEN PRINT “Response = ” LEFT$(MSG$,
LFPOS)
480
IF LFPOS = 0 THEN PRINT “Response = ”; MSG$
490
CALL IBRSP (CNT%, SPR%)
500
WEND
510
520
‘ —— CHECK EVENT STATUS BIT ——
530
IF NOT SPR% AND 32 GOTO 750
540
PRINT “ Event status bit set”
550
WRT$ = “*esr?”
560
CALL IBWRT (CNT%, WRT$)
570
ESR$ = SPACE$(255)
580
CALL IBRSP (CNT%, SPR%)
590
CALL IBRD (CNT%, ESR$)
600
ESR% = VAL(ESR$)
610
IF ESR% AND 32 THEN PRINT “
Command error”
620
IF ESR% AND 16 THEN PRINT “
Execution error”
630
IF ESR% AND 4 THEN PRINT “
Query error”
640
650
‘ —— READ ERROR MESSAGES ——
660
WRT$ = “syst:err?”
670
ERRMESS$ = SPACE$(255)
680
CALL IBWRT (CNT%, WRT$)
690
CALL IBRD (CNT%, ERRMESS$)
700
WHILE NOT INSTR(ERRMESS$, “No error”) <> 0
710
PRINT LEFT$(ERRMESS$, INSTR(ERRMESS$, CHR$(10)))
720
CALL IBWRT (CNT%, WRT$)
730
CALL IBRD (CNT%, ERRMESS$)
740
WEND
750 WEND
760 PRINT “PROGRAM TERMINATED”
770 END
Programming Examples
4-10 GW-Basic for National Instruments PC-IIA, Setting Up the Interface