OPTIONAL HARDWARE and SOFTWARE 
Teledyne API Model 701 Zero Air Generator 
4-2 
01671H (DCN6051) 
 
CAUTION 
THE AIR EXITING THE SCRUBBER IS AT 300
C (575F). THIS HOT 
AIR ENTERS THE COPPER TUBING WHERE IT IS COOLED BEFORE 
IT ENTERS THE NO SCRUBBER. 
THIS COPPER TUBING AND THE CASING CAN BE VERY HOT. 
 
The catalyst is proprietary. It theoretically should never need changing. However, it may 
become contaminated or poisoned over time so we recommend that the scrubber be 
replaced once per year or sooner. 
4.3 Carrying Strap/Handle (OPT 29) 
The chassis allows the user to attach a strap handle for carrying the instrument.  The 
handle is located on the right side and pulls out to accommodate a hand for transport.  
When pushed in, the handle is nearly flush with the chassis, only protruding out about 9 
mm (3/8”).   
 
Figure 4-1:  Carrying Strap Handle and Rack Mount Brackets 
 
Installing the strap handle prevents the use of the rack mount slides, although the rack 
mount brackets, OPT 21 can still be used.  
 
CAUTION
 
 
G
ENERAL 
S
AFETY 
H
AZARD
 
A fully configured M701 with valve options weighs about 55 lbs (25 kg). To 
avoid personal injury, use  two personsto  lift and carry the instrument. 
Ensure to disconnect all cables and tubing from the analyzer before 
moving it.