5.12 Pressure Gauge 
The front panel-mounted pressure gauge shows approximately the regulated air 
pressure available to a calibrator. The gauge will usually indicate 30 psig. If the flow is 
very low, as in a standby condition, the gauge may read a little higher, say 33 psig. It 
will move to the correct reading when the air flow is increased. If the pressure drops 
below 30 psig, it may be because too much air is being commanded. At maximum air 
flow, 20 SLPM, you may notice that the pressure drops to 29 psig. This is normal. 
5.13 Hydrocarbon Scrubber (option) 
The optional Hydrocarbon scrubber catalytically converts hydrocarbons to carbon 
dioxide and water. The scrubber consists of a cylindrical cartridge which contains a 
preheat chamber and the pelletized precious-metal catalyst. The canister is heated by a 
band heater with an integral type-K thermocouple and is housed in a stainless steel 
casing with appropriate thermal insulation. The cartridge temperature is controlled by 
the M701 control PCB. The controller reads the thermocouple voltage, compensates for 
cold-junction temperature and modulates the heater to maintain a constant cartridge 
temperature. (See Appendix, drawing 00257). 
The operating temperature of the cartridge is 300ºC. The temperature can be verified by 
measuring the thermocouple voltage at J12, the terminal block at the edge of the board 
nearest the Hydrocarbon Scrubber. The positive pin is clearly marked. At 300ºC, the 
thermocouple voltage is 11.2 mV in an ambient temperature of 25ºC. (11.4 mV at 20º 
and 11.0 mV at 30º). While the controller compensates for cold-junction temperature in 
controlling the heater, the thermocouple voltage at the terminal block is not 
compensated, so allowance must be made for ambient temperature when converting 
the voltage to cartridge temperature. Pot R3, near the top right corner of the PCB, 
controls the temperature set-point. (Clockwise to increase.) LED DS1 blinks when the 
temperature is at the set-point. LED DS2 is an alarm and will be ON if the thermocouple 
is OPEN. Pot R4 has no effect in this application. 
 
 
CAUTION 
The air exiting the scrubber is at 300
C (575F). This hot air is 
cooled in the copper tubing before it enters the NO scrubber. 
This coil and the casing can be very hot. 
 
The catalyst is proprietary. It theoretically should never need changing. However, it may 
become contaminated or poisoned over time so we recommend that the scrubber be 
replaced once per year or sooner.