30    Basic setups 
 
  Measurement setup (for community noise) 
 
 
Models SE and DL 
C-A measurement settings 
The C-A measurement is determined from concurrent measurements of the 
SPL by both meters. It can only be made if certain settings for Meter 1 and 
Meter 2 are compatible. If any incompatibility exists, the L
c-a
 selection in this 
screen will be marked N/A (Not Available).  
 
If you want to activate L
c-a
, have the following settings in place before you 
visit the Measures screen.  
• 
Thresholds - Both meters set to “Off” or the same value.  
•  
Exchange rates- Both meters set to the same value.  
•  
Time response - Both meters set to the same value. 
•  
Frequency weightings- Meter 1 must be set to A-weighting; Meter 2 
must be set to C-weighting.  
Figure 3-6 is an example of the appropriate C-A settings for Meter 1 (left 
side) and Meter 2 (right side). (Please see “Meter 1/Meter 2 Parameters” on 
page 27 for setup details.) 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 3-6:  C-A measurement settings for Meter 1 and Meter 2 
Meter 1  
• 
Press softkey to 
toggle between 
Meter 1/Meter 2 
Slow response  A-weighting  
 Slow response  C-weighting   Meter 2