Description and Specifications
Integral Cable 396P: Code 01 - 25 ft; Code 02 - 15 ft coaxial / 396PVP: none - must use mating
VP cable
Preamplifier Options: Remote or Integral SMART preamplifier
Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/2 lb)
Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/
1.3
Ordering Information
The 396P Sensor is housed in a molded reinforced polypropy-
lene body with 1 in. MNPT threads suitable for insertion,
submersion or flow through installation. The sensor includes a
general purpose pH electrode or a platinum ORP electrode, a
reference junction and a solution ground. The 396P comes
standard with a recessed electrode; an optional slotted tip is
also available. In addition, the 396P features an optional inte-
gral hermetically sealed preamplifier and 15 ft or 25 ft cable
lengths. Automatic temperature compensation,  Pt 100 or 3K
Balco, is standard with the 396P.
pH Range 
396 / 396VP
396P / 396PVP
396R / 396RVP
GPHT Hemi
GPHT Flat
GPHT Hemi
GPLR Flat
GPHT Hemi
GPHT Flat
0-2 pH 
94%
95%
94%
94%
93%
2-12 pH 
99%
99%
97%
98%
97%
98%
12-13 pH 
97%
96%
98%
95%
98%
95%
13-14 pH 
92%
98%
98%
*Percent Linearity
CODE
PREAMPLIFIER/CABLE (Required Selection)
01
With integral SMART preamplifier, 25 ft (7.6m) cable
02
Without integral preamplifier, 15 ft (4.6m) cable
CODE
MEASURING ELECTRODE TYPE (Required Selection)
10
GPLR hemi bulb, General Purpose Low Resistivity (0-14 pH) 
12
ORP
13
GPLR flat bulb, General Purpose Low Resistivity (2 - 13 pH)
CODE
ANALYZER/TC COMPATIBILITY (Required Selection)
50
For 1181 (3K TC)
54
For 1054A/B, 2054, 2081 (Pt 100 RTD)
55
For 54e, 3081, 81, 4081, 1055, 5081, Xmt (PT-100 RTD)
396P
01
-
10
-
55
EXAMPLE
NOTE: The 396P is also compatible with Model SCL-P/Q (option 02-54 only).
CODE
OPTIONAL SELECTION
41
Slotted Tip (not available on flat bulb sensors)