P851 Short Shot Floor
Enter the minimum echo strength (in dB above 1 μV rms) derived from a short shot to be 
considered for evaluation.
P852 Short Shot Range
Enter the maximum distance in Units (P005) to be measured using short shot echoes.
This feature is automatically altered when Transducer (P004) is altered. 
Test (P900 to P913)
P900 Software Revision Number
View the EPROM Rev. #.
P901 Memory
Press ENTER 
 to activate the SITRANS LU01 memory test.
Values
Range: 30 to 100
Preset: 50
Related
P840 Short Shot Number
P842 Short Shot Frequency
P844 Short Shot Width
P850 Short Shot Bias
P852 Short Shot Range
Values
Range: 0.000 to 9999
Altered By
P004  Transducer
Related
P840 Short Shot Number
P842 Short Shot Frequency
P844 Short Shot Width
P850 Short Shot Bias
P851 Short Shot Floor
Note: 
These parameters are for Siemens Milltronics service personnel only.
Values
Range: 00.00 to 99.99 (view only)
Values
Display: view only
PASS
(memory test successful)
F1
RAM
F2
NOVRAM
F3
EEPROM
F4
EPROM