National Instruments NI 4070/4072 6 1/2-Digit FlexDMM Manuel D’Utilisation

Page de 68
© National Instruments Corporation
35
NI 4070/4072 Calibration Procedure
To verify the frequency of the NI 4070/4072, complete the following steps:
1.
Remove all connections from the NI 4070/4072.
Note
Polarity is not important in steps 2, 3, and 5.
2.
Connect one end of the coaxial cable to the Pomona 4892 double 
banana plug.
3.
Tighten the other end of the coaxial cable in the screw terminal 
channels 5 and 39 of the TB-2715 terminal block.
4.
Connect the TB-2715 with the coaxial cable attached to the NI 6608.
5.
Plug the Pomona 4892 into the HI and LO terminals of the 
NI 4070/4072.
6.
Call 
niDMM_reset
 to reset the NI 4070/4072 to a known state. 
7.
Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_FREQ
Range = 
1
Resolution = 
0
8.
Call 
niDMM_ConfigureFrequencyVoltageRange
 with 
Voltage Range set to 
5
.
9.
Call 
GPCTR_Control
 with the following parameters:
deviceNumber = the device number of the NI 6608, assigned by 
Measurement & Automation Explorer (MAX)
gpctrNum = 
ND_COUNTER_0
action = 
ND_RESET
10. Call 
GPCTR_Set_Application
 with the following parameters:
deviceNumber = the device number of the NI 6608, assigned 
by MAX
gpctrNum = 
ND_COUNTER_0
application = 
ND_PULSE_TRAIN_GNR
11. Call 
GPCTR_Change_Parameter
 with the following parameters:
deviceNumber = the device number of the NI 6608, assigned 
by MAX
gpctrNum = 
ND_COUNTER_0
paramID = 
ND_COUNT_1
paramValue = 
10e6