Chapter 2
Configuration and Installation
© National Instruments Corporation
2-5
W38 in its factory default position (position 1). On all but one of the 
SCXI-1121s that are cabled to the data acquisition board, move jumper 
W32 to position 3. It does not matter which of the SCXI-1121 modules that 
are cabled to the data acquisition board has jumper W32 set to position 1. 
If you have different types of modules cabled to the data acquisition board, 
those different modules will have jumpers similar to W38 and W32 of the 
SCXI-1121. Set those jumpers on the different modules using the same 
method described here for the SCXI-1121.
On Revision A and B SCXI-1121s, jumper W32 is not used. You set jumper 
W38 as explained in the cases above, except in the case of a multiple 
chassis ribbon cable system. In a multichassis ribbon cable system with 
Revision A and B SCXI-1121s cabled to the data acquisition board, you can 
access the MISO line in only one chassis. Pick one of the chassis and set 
jumper W38 to position 1 on the SCXI-1121 in that chassis that is cabled 
to the data acquisition board. On the SCXI-1121s that are in the other 
chassis and cabled to the data acquisition board, set jumper W38 to 
position 3. Notice that you will only be able to access digital information 
from the chassis that has the SCXI-1121 with jumper W38 set to position 1.
Table 2-1.  Digital Signal Connections, Jumper Settings
Jumper
Description
Configuration
W38
Factory setting; 
connects MISO to 
SERDATOUT
W38
Parking position
W45
Factory setting
3
2
1
3
2
1
3
2
1