IP Protection
Cleaning instrucions
IPX0 
Clean with a damp cloth
IPX1 
Clean with a damp cloth
IPX2 
Clean with a damp cloth
IPX3 
Clean with a damp cloth
IPX4 
Clean with a damp cloth
IPX5 
Wash with a brush and water, but not water under pressure
IPX6 
X Wash with a brush and water. The water can be under pressure, but the system 
must not be hosed down directly with a high-pressure cleaner. Max. 20oC
IPX6 Washable
Clean by the use of water tunnels 
IP Protection degree:
All LINAK products can be cleaned as follow according to their IP protection, which 
is stated on the product label.
The IP code specifi es the degrees of protection provided by the enclosures. Only 
the protection against ingress of water (second characteristic numeral) is specifi ed, 
ingress of solid foreign objects (fi rst characteristic numeral) is not specifi ed and 
therefore replaced by the letter X in the code.
The materials are resistant to the majority of cleaners and disinfectants used in the 
hospital and nursing home sector.
Cleaning with a steam cleaner, for example, is not permitted, as it will not be 
possible to maintain a minimum distance of 300 mm from the electrical parts.
The following directions regarding cleaners and disinfectants must be 
complied with:
•  They must not be highly alkaline or acidic (pH value 6-8).
•  They must not contain caustic agents.
•  Their contents must not be able to change the structure of the surface or
  adhesion of the plastic.
•  They must not break down grease.
The following disinfectants/detergents are recommended:
•  Neodischer Dekonta, dosage 1-3%, made by CHEMISCHE FABRIK Dr. WEIGERT,
•  CCOTRADE RW, dosage 0.5%, made by CCOTRADE GmbH.