Begin
2.2.4
Determine your connection type
Different transmitters and networks support different connection types. 
Table 2-2 lists the supported protocols and wiring methods for IFT97xx and RFT97xx 
transmitters.
Table 2-3 lists the supported protocols and wiring methods for Model 1500/2500, 
Model 1700/2700, and Series 3000 transmitters.
Table 2-4 lists the supported protocols and wiring methods for the Model 2400S transmitter.
Table 2-5 lists the supported protocols and wiring methods for all MVD Direct Connect 
systems.
In these tables, and throughout the chapter:
Temporary – refers to a connection that is not permanent and is typically used only for 
configuration and troubleshooting. Because the transmitter housing must be open for the 
duration of the connection, these connections should be removed and the housing closed as 
soon as possible. The operator should be aware of the safety hazards that result from opening 
the transmitter housing.
Hard-wired – refers to a connection that is made to the permanent wiring, usually a transmitter 
output wire or the network that the transmitter is already using. Because hard-wired 
connections do not require the transmitter housing to be open, they can be left in place as 
desired.
AN – refers to transmitters with the analog outputs option board
IS – refers to transmitters with the intrinsically safe outputs option board
CIO – refers to transmitters with the configurable input/outputs option board
FF – refers to transmitters with the F
OUNDATION
 fieldbus input/output option board
PA – refers to transmitters with the PROFIBUS-PA input/output option board
DP – refers to transmitters with the PROFIBUS-DP input/output option board
DN – refers to transmitters with the DeviceNet input/output option board
MVD Direct Connect – refers to meter installations that include the core processor but do not 
include a transmitter. ProLink II is connected directly to the RS-485 terminals on the core 
processor or the MVD Direct Connect I.S. barrier.
Once you have determined your connection type, use the Wiring Method # value in the table to direct 
you to the correct wiring procedure in Section 2.2.5.