A&D AD-4212B-101/102/201/301 Manuel D’Utilisation

Page de 102
 
45
Note 
The upper/lower limit values can be confirmed by external commands using the RS-232C serial 
interface. 
 e.g., 
 
Command 
?HH 
  (to 
confirm secondary upper limit value) 
  
Response 
HH,  
+100.00000 g 
(secondary upper limit value=100.00000 g) 
   
Command 
?HI   
 
(to confirm upper limit value) 
  
Response 
HI +080.00000 g (upper limit value=80.00000 g) 
   
Command 
?LO 
 
(to confirm lower limit value) 
  
Response 
LO,  
+060.00000 
(lower limit value=60.00000 g) 
  
Command 
?LL 
  (to 
confirm 
secondary lower limit value) 
  
Response 
LL, 
+040.00000 g 
(secondary lower limit value=40.00000 g) 
   
 
                  indicates a space (20h) 
 
Adding the comparison results (AD-4212A only) 
To add the comparison results to the output data, set “Comparison results (Cp-r)” parameter of the 
function table to “1”. 
 
ST, H H , 110.0000           g 
HH (Only when OP-04 is installed.) 
 
ST, HI,  +090.0000      g 
HI 
 
ST, OK, +070.0000     g 
OK  
 
ST, LO, +050.0000      g 
LO  
 
ST, LL,+030.0000          g 
LL  (Only when OP-04 is installed.) 
 
ST, - -,+000.0000      g 
No comparison when unstable or near zero   
   
 
   
       indicates a space (20h) 
 
Notes 
Select A&D standard format. Set “Data format (
type)” parameter of the function table to “0”. 
The comparison results can not be added to the weighing data stored in memory.