A&D AD-4212B-101/102/201/301 Manuel D’Utilisation
45
Note
The upper/lower limit values can be confirmed by external commands using the RS-232C serial
interface.
e.g.,
Command
?HH
(to
confirm secondary upper limit value)
Response
HH,
+100.00000 g
(secondary upper limit value=100.00000 g)
Command
?HI
(to confirm upper limit value)
Response
HI +080.00000 g (upper limit value=80.00000 g)
Command
?LO
(to confirm lower limit value)
Response
LO,
+060.00000
g
(lower limit value=60.00000 g)
Command
?LL
(to
confirm
secondary lower limit value)
Response
LL,
+040.00000 g
(secondary lower limit value=40.00000 g)
indicates a space (20h)
Adding the comparison results (AD-4212A only)
To add the comparison results to the output data, set “Comparison results (Cp-r)” parameter of the
function table to “1”.
ST, H H , 110.0000 g
HH (Only when OP-04 is installed.)
ST, HI, +090.0000 g
HI
ST, OK, +070.0000 g
OK
ST, LO, +050.0000 g
LO
ST, LL,+030.0000 g
LL (Only when OP-04 is installed.)
ST, - -,+000.0000 g
No comparison when unstable or near zero
indicates a space (20h)
Notes
Select A&D standard format. Set “Data format (type)” parameter of the function table to “0”.
Select A&D standard format. Set “Data format (type)” parameter of the function table to “0”.
The comparison results can not be added to the weighing data stored in memory.