A&D AD-4212A-100/200/600/1000 Manuel D’Utilisation
19
6-2 Smart Range Function
The AD-4212B-101 is equipped with two ranges. The precision range has a higher resolution. The
standard range has normal resolution.
The range switches automatically, depending on the value displayed. Pressing the RE-ZERO key
allows weighing in the precision range, regardless of the tare value.
The minimum weighing value can be fixed to 0.1 mg or 1 mg by pressing the SAMPLE key.
1 Press the RE-ZERO key.
The balance will start weighing, using the
The balance will start weighing, using the
precision range.
2 Place a container on the weighing pan.
The weight value displayed exceeds the
The weight value displayed exceeds the
precision range value and the balance will
switch to the standard range.
switch to the standard range.
3 Press the RE-ZERO key.
The balance will switch to the precision range.
The balance will switch to the precision range.
4 Place a sample in the container.
The weight value displayed is within the
The weight value displayed is within the
precision range value and the balance will
perform a weighing, using the precision range.
perform a weighing, using the precision range.
Precision range/standard range value
Weighing range
Available minimum weighing value
Precision range
0 g to 31 g
0.01 mg 0.1 mg 1 mg
Standard range
31 g to 110 g
0.1 mg 1 mg
* The factory setting of the minimum weighing value is 0.1 mg (0.0001 g).
Standard range display
Precision range display
Precision range display
Precision range display
Container
Weighing pan
Sample