Siemens 7ML19985AB01 Manuel D’Utilisation

Page de 112
Modem Communication
Communication with the OCM-3 over an answer back telephone modem is relatively
straight forward. For modem communication over conventional telephone lines, a
maximum of 1200 or 2400 baud is recommended. Generally, a lower baud rate is
more reliable on noisy lines. Modem should be self answering and data checking
capability turn on.
The Milltronics Utilities software can be run via modem. In addition the Terminal
Emulation program of the Utilities software package can be used to send control
sequences directly to the modem.
The remote modem is defined as the modem connected to the OCM-3. The local
modem is defined as the modem connected to the computer.
It should be noted that the modem responses listed in the following discussion serve
only to typify what is presently common in current field of RS-232 communication.
OCM - 3
C
O
M
R
X
D
T
X
D
RS 232
I/O
11
2
3
7
OCM-3
TB-1
12 13
remote modem
local modem
computer
standard modem
cable
telephone  lines
modem
DB-25 connector
7ML19985AB01
OCM III
110