Considerations
- set the ’cold start’ (F12) default baud to 1200 by removing jumper J1 on board A.
- set both modems and the OCM-3 baud rates to 1200. 
- set the remote modem to operate in the ’answer back’ mode. 
  Refer to modem’s user manual.
Communication
- select serial programming link with OCM-3, from the utilities menu. See Milltronics
  Utilities Software Disk.
The computer screen will display the letters ’v’. These would normally 
have gone to the OCM-3 to ask it to send its display, but the modem 
will have returned it to the screen instead.
Establishing Communication
- type the following:
ATD
  and the telephone number of the remote modem, then press ‘ENTER’.
The modem will dial the number you have entered. When the local
modem establishes contact with the remote modem, the local modem 
sends a message to the screen (typical).
CONNECT
- press the ‘v’ and ‘ENTER’ keys to turn on the OCM-3’s display-to-serial channel.
 
The OCM-3 is now on line and communication can be carried out. The user 
may escape to the main menu of the Utilities software and select another 
program if so desired.
 
To close the communication, return to the Terminal Emulation program.
7ML19985AB01
OCM III
111